This article refers to the workings of Smith’s Medical Pneupac ParaPac and the newer ParaPac Plus. Other transport ventilators may vary in their settings and processes. Please check what unit your ambulance trust has before following the information here.
[tabby title=”Detail”]
The pressure of air inside our lungs is normally just right to not only ensure that negative pressure breathing occurs but also to prevent a shunt (where blood bypasses the lungs) (Hess & Kackmarek, 2014: 431), the oesophageal spinchter from opening, causing air to enter the stomach (Baker, 2012: 1), and a host of other potential problems. Any means of supported ventilation where the patient is not breathing for themselves produces positive pressure ventilation. Not only does this change the way air enters the patient’s lungs while they are ventilated, but it also comes with all of the above risks if any of the settings are incorrect, and more.
Many of the effects ‘are related to mean airway pressure [which is] […] the average pressure applied to the airway during the ventilatory cycle’ (Hess & Kackmarek, 2014: 420) With too ⇑ pressure comes ⇓ venous return (Hess & Kackmarek, 2014: 420) and ⇑ chance of barotrauma (Baker, 2016: 75).
To summarise, the potential negative effects are:
Blood bypassing the lungs (a shunt) (Hess & Kackmarek, 2014: 431)
Because modern ventilators are volume-controlled, preventing a different problem that occurs during pressure-controlled ventilation, there is a set tidal volume which the ventilator will always provide to the patient. To provide this volume, the pressure will constantly increase until the volume is reached (Baker, 2016: 132).
The Means of Avoiding These Problems
In ventilators, the pressure of an oxygen cylinder ‘is lowered through a reducing valve down to 30-90psi’ from 2000psi. It’s then delivered to the patient ‘to a maximum of 40cmh20 or 60cmh20’, (Baker, 2012: 18)*, depending on whether you’re in the UK or the US.
*cmh20 = centimetres of water pressure (SensorsONE, 2017)
Since we can’t have too much pressure there needs to be a way of stopping the ventilator if things get awry. The relief pressure is what this refers to. Via a safety valve (Baker, 2012), otherwise known as ‘exhaust valves’ the supply is cut and excess peak inspiratory pressure (PIP) is released into the atmosphere (Zaconeta et al, 2010), ‘prevent[ing] over-ventilation of the lungs’ (Baker, 2016: 145)
Why would the pressure get this high anyway?
‘The patient circuit […] [could be] blocked or kinked’ (Baker, 2016: 142). Best to check it.
‘If the lung compliance[*] is low and the airway resistance is high the pressure can rise to high levels’ because, as above, the pressure will keep rising until the set tidal volume is delivered (Baker, 2016: 142).
The wrong relief pressure settings.
*Compliance ‘refers to the ease of expansion of the lungs and the walls of the thorax’ (Baker, 2012: 8). Ordinarily, according to Baker (2012: 8), the value is 100 cmh20 but when ventilated it’s 40-50cmh20. Physical injuries and medical conditions may further alter this figure.
Which dial are you talking about?
This one
The Pneupac ParaPac (Smiths Medical). Photo highlights the Relief Pressure dial.
What should I set it to?
40cmh20 if you’re in the UK (Baker, 2012)
Why is the relief pressure set at 40cmh20 in the UK? Why is it 60cmh20 in the US?
Article 999 is struggling to find the answers to these questions. If you know the answers, please get it touch. It would be great to add the information to this resource.
This post is yet to be peer reviewed. Please get in touch if you have any comments.
[tabby title=”Super Summary”]
Article 999 previously posted an article on positive pressure ventilation, including its risks. The aim of positive pressure ventilation is to ‘maximise the potential benefit of mechanical ventilation while minimizing the potential for harm’ (Hess & Kackmarek, 2014: 407). The relief pressure dial on the ventilator ensures that the pressure of ventilation is not so high as to cause blood bypassing the lungs (Hess & Kackmarek, 2014: 407), the oesophageal spinchter opening (Baker, 2012: 1), barotrauma (Baker, 2016: 75; Hess & Kackmarek, 2014) and more. The ventilator’s other settings ensure that the patient actually receives ventilation, but the relief pressure setting ensures that a valve drains excess air when settings, poor setup, or the patient’s lung compliance (Baker, 2016: 43) have caused the pressure to get to high.
What should I set it to?
40cmh20 (Baker, 2012)
What should I look out for if the pressure is too high?
Kinked tubing (Baker, 2016: 142) & your settings
Read more in the Detail tab.
[tabby title=”Prefer to see a diagram? Check out the mind-map summary”]
[tabby title=”References”]
References
Baker, D. 2012. Emergency and Transport Ventilation, an introductory guide, Smiths Medical International Limited: Luton
Hess, D.R. & Kacmarek, R.M. 2014. Essentials of Mechanical Ventilation. McGraw Hill Education: New York
SensorsONE, 2017. cmh20 – Centimetres of Water Column at 4 deg C Pressure Unit, Available Online: https://www.sensorsone.com/cmh2o-centimetres-water-column-4-deg-c-pressure-unit/ (Accessed 16/11/17)
Zaconeta, C.A.M. et al, 2010. ‘Evaluation of peak inspiratory pressure and respiratory rate during ventilation of a preterm infant lung model with a self-inflating bag by paramedics of the Fire Department’, Revista Paulista de Pediatria, vol. 28. No. 1.
Not endorsed by but with thanks to Smiths Medical for providing information.
Positive pressure refers to the way artificial ventilation ‘inflate[s] the lungs’ (Hess & Kacmarek, 2014). ‘Normal breathing depends on drawing in air to the lungs by creating a partial vacuum inside the chest cavity’ (Baker, 2016: vi). The mean airway pressure, an average combined of ‘the amount and duration of pressure applied during’ expiratory and inspiratory phases in normal breathing is negative (Hess & Kacmarek, 2014: 420 ebook). See the article on negative pressure breathing for more information. Intermittent positive pressure ventilation, on the other hand, is for inspiration the ‘exact reverse of this process with air being forced into the lungs’ (Baker, 2016: vi).
Positive pressure ventilation includes the use of bag-valve-masks, but the majority of this article will focus on the ventilator.
Upcoming articles will discuss the specific settings and how-tos relating to the Pneupac ParaPac ventilator. Current links:
Ventilators have 4 phases of operation:
1. ‘The inspiratory phase’
2. ‘Cycling between the inspiratory (I) and expiratory (E) phase’
3. ‘The expiratory phase’
4. ‘Cycling between the E and I phase’
(Baker, 2016: 117)
The majority of ventilators are set to an I (Inspiratory) to E (Expiratory) ratio of 1:2, i.e. the expiratory phase is twice as long as the inspiratory phase, ‘which mirrors normal breathing’ (Baker, 2016: 108). If required, some ventilators allow this setting to ‘be altered to assist with restrictive lung diseases’ (Baker, 2016: 108). ICU ventilators contain additional settings that alter this ratio, including a ‘respiratory pause’ between the I and E phases ‘to allow distribution of gas within the lungs’ (Baker, 2016: 108).
[tabby title=”Mechanics”]
Mechanics
So, inspiration occurs when air is ‘forced into the lungs’ (Baker, 2016: vi). Expiration, however, occurs when ‘the ventilator stops delivering a positive pressure and the patient breaths out passively, as in normal breathing. This passive expiration (E) is ensured by the elastic recoil of the lungs and the chest wall’ (Baker, 2016: 108). The timing is aided by a ‘pneumatic oscillator’ (Baker, 2016: 22).
The gas is delivered by the ‘patient circuit’, which also ‘evacuate[s] expired CO2’ (Baker, 2016: 109). This is delivered ‘as close as possible to the airway device being used’ by a valve (Baker, 2016: 109). There are also filters to prevent contamination of the ventilator from the patient.
Early ventilators had a ‘lack of interactivity with the patient’s own breathing efforts if there was only partial respiratory failure’ (Baker, 2016: 22) but modern ventilators tend to have different modes to account for this and additional settings beyond a simple tidal volume dial. In particular, ventilators feature a ‘demand valve’ which detects when a patient takes a breath (Baker, 2016: 22). There are also settings for tidal volume and ventilation frequency. These settings will be discussed in upcoming articles.
The main positive effect of artificial ventilation is fairly obvious: it assists with the ventilation of a patient who is completely or partially unable to breathe normally. There are however a number of negative effects created by positive pressure ventilation. Related to these is the fact that mechanical ventilation can cause ‘heterogeneity’ of the lungs, meaning that ‘some lung units are prone to overdistension and others are prone to collapse’ (Hess & Kacmarek, 2014: 493). These are part of the reason most ventilators have relief pressure settings. This setting will be discussed in an upcoming article. You can check back here for the link.
Positive pressure ventilation can cause or effect the following:
The following are unavoidable results of ventilator use:
Decreased cardiac output
…Leading to ‘hypotension and potential tissue hypoxia. This effect is greatest with high mean airway pressure, high lung compliance, and low circulating blood volume’ (Hess & Kacmarek, 2014: 513). The high pressure ‘decreases venous return and right heart filling, which may reduce cardiac output’ (Hess & Kacmarek, 2014: 513).
This pressure is normally lowest ‘during inhalation’, so venous return would be highest. But with positive pressure ventilation, ‘venous return is greatest during exhalation’ as opposed to inhalation (Hess & Kacmarek, 2014: 513-520).
Renal and gastric effects
Mechanical ventilation reduces urine output due to ‘decreased renal perfusion’ which results from ‘decreased cardiac output’ (Hess & Kacmarek, 2014: 527). Unfortunately this can cause many of the issues common with fluid retention. Similarly, gastric distension can result, along with gastrointestinal bleeding’ (Hess & Kacmarek, 2014: 533).
Increased intracranial pressure
…in head injury patients, due to ‘a decrease in venous return, which increases intracranial blood volume and pressure’ (Hess & Kacmarek, 2014: 540).
Pneumonia may occur, especially due to invasive ventilation which is the form in emergency practice (non-invasive refers to ventilatory support such as CPAP). It ‘results from aspiration of oropharnygeal secretions around the cuff of the endotracheal tube’ (Hess & Kacmarek, 2014: 499).
The following can occur as a result of ventilator settings:
Hyperventilation and hypoventilation
These risk effecting the pH, and causing respiratory alkalosis (Hess & Kacmarek, 2014: 499)
Oxygen Toxicity due to high oxygen levels (Hess, 2014: 506)
A Shunt is usually partly decreased but settings can change this:
A shunt is ‘perfusion (blood flow) without ventilation’, which happens ‘when blood flows from the right heart to the left heart without participating in gas exchange’ (Hess & Kacmarek, 2014: 427). There are two different types of shunts: capillary and anatomic. The first occurs when the alveoli are not ventilated. One example of this is pneumonia (Hess & Kacmarek, 2014: 427). The latter type of shunt occurs when blood flow ‘completely bypasses the lungs’, which can sometimes happen with ‘congenital cardiac defects’ (Hess & Kacmarek, 2014: 427).
Positive Pressure Ventilation’s typical effect is to decrease capillary shunt by improving ‘the distribution of ventilation, particularly by improving the ventilation of previously underventilated areas of the lungs’ (Hess & Kacmarek, 2014: 427-446)
Despite the above improvements, if the settings are not ideal hypoxemia can result. This is the same outcome as a shunt but would in this case be caused by ‘redistribution of pulmonary blood flow to unventilated regions’ due to ‘overdistension of some lung units’ (Hess & Kacmarek, 2014: 324). Positive pressure ventilation can also increase anatomic shunt (Hess & Kacmarek, 2014: 441)
Also due to incorrect settings, the following can occur as a lack of pressure or volume:
Atelectasis occurs when the pressure or volume during or after artificial ventilation isn’t enough to inflate the lungs –> complete or partial collapse.
Atelectasis ‘is larger with obese patients and when a higher fraction of inspired oxygen (FI02) is used […]. During mechanical ventilation […] atelectasis may occur when lungs are underinflated due to low tidal volumes, or when compression occurs (such as patient position or obesity)’ (Open Anesthesia, 2017). If the volume or pressure isn’t enough to ‘prevent the alveoli from collapsing’ (Open Anesthesia, 2017) then regardless of ventilator type (pressure vs volume controlled) atelectasis may result.
The following can occur as a result of apparatus setup:
Dead Space can be increased by mechanical ventilation (Hess & Kacmarek, 2014: 471). It may be ‘reduced by an artificial airway (ETT) but is increased by increasing distance from the end of the ETT to the means of evacuating CO2 in mechanical ventilation’ (Baker, 2016: 111). Size, age, upright postures and neck extension also increase it (Baker, 2016: 111).
The following can occur due to overdistension:
Barotrauma = ‘alveolar rupture due to overdistension’, which can cause numerous problems including pneumothorax (Hess, 2014: 480). Volume generators have a ‘safety valve which activates at a preset pressure’ to avoid barotrauma’ (Baker, 2016: 118). This will be discussed in an upcoming article.
Ventilator-Induced lung injury caused by overdistension. Massively effected by ventilator settings for tidal volume and relief pressure (Hess, 2014: 493)
There are additional risks of positive pressure ventilation, particularly mechanical ventilation. Some of these are related to the use of airway adjuncts. Others are related to the removal of the ventilator. To read more, check out the references below.
[tabby title=”Types of Positive Pressure Ventilation”]
Types of Positive Pressure Ventilation
‘A spectrum of modes exist between total artificial ventilation and a patient who is still breathing spontaneously but with reduced efforts’ (Baker, 2016: 105).
There are two main types of ventilators:
– Pressure generated
– Volume generated
Pressure generators ‘provide a constant pressure during the inspiratory phase’ (Baker, 2016: 118). Volume generators release gas once a set volume has been reached. In each one, only the pressure or the volume is a constant. Everything else is relative and will adjust until the preset parameter of volume or pressure has been reached (Baker, 2016; Hess, 2014).
Issues
‘The problem of pressure generation is that it cannot guarantee an adequate Vt [volume] in situations where the lung mechanics are abnormal or where a guaranteed Vt is required due to life-threatening hypoxia. They are therefore not suitable for most emergency ventilation, with the notable exception of neonatal and infant ventilation where flow generators are contraindicated due to the fragility of the lung tissue’ (Baker, 2016: 118).
Additional Types
Most ventilators switch from inspiration to expiration by ‘time cycling’ (Baker, 2016: 119). However, this can also be due to volume or pressure cycling. The time cycling ratio is usually 1:2 for I:E, in other words the expiration time is twice as long as the inhalation time. ‘Ventilation modes can best be understood as a spectrum describing the work of breathing done by the ventilator and by the patient’ (Baker, 2016: 122.
The modes are: Controlled Mandatory Ventilation, in which ‘ventilation is provided entirely by the ventilator’ (Baker, 2016: 122), Assisted Controlled Ventilation which is fairly self-explanatory, and Synchronised Mandatory Ventilation, in which ‘the patient’s respiratory efforts in initiating a ventilation are synchronised with ventilations provided automatically by the ventilator.’ This is best used to supplement the patient’s efforts. It’s otherwise known as ‘assist-control’ and is typically found in ICU. (Baker, 2016: 123).
In portable ventilators the modes are:
Intermittent Mandatory Ventilation, Synchronised intermittent mandatory ventilation, and synchronised mandatory minute volume but only the latter two are still in use. The second overcomes the problems of the first, ensuring that breaths are supplemented by the ventilator and that supplemented breaths are activated only when there is a certain trigger e.g. low pressure. There is a modern, more complex version of this now in use in hospitals but this may exist in emergency ventilation practice to assist a patient who is ‘in partial respiratory failure’ (Baker, 2016: 124).
The first two divide minute volume between the ventilator and the second. The latter is based on time, which is compared with a ‘set minute volume. If there is a discrepancy between the actual minute volume and the set delivered minute volume, an automatic ventilation is delivered which corrects the difference’ (Baker, 2016: 124).
Because there is an issue of the ventilator being unable to tell the difference between dead space ventilation and alveolar ventilation, Pneupac have modified the above with a series of Ifs (if the patient takes a breath, if there is no further breath, if the tidal volume is too low, etc) that adjusts settings accordingly (Baker, 2016: 124). There are more modes that exist for ICU ventilators, which tend to be a lot more complex in design (Baker, 2016).
CPAP is also a form of positive pressure breathing and is designed ‘to improve oxygenation in lungs that may not be exchanging oxygen with the blood adequately’ (Baker, 2016: 126). There is a constant pressure in CPAP, versus ‘different pressures during inspiration and expiration’ in BIPAP’ but both are designed to allow the patient to do all the work of breathing. They merely increase oxygen and make inspiration easier while increasing the work of expiration. ‘Breathing with CPAP has been compared to the sensation of breathing against a very strong wind’ (Baker, 2016: 128). They ‘open up (recruit) alveoli in the lungs that may be non-ventilated.’ (Baker, 2016: 127). Both push back fluid ‘into the circulation via lymphatic drainage of the lung interstitial space’ and increase ‘the diffusion of oxygen into the pulmonary capillaries’ (Baker, 2016: 127).
[tabby title=”PEEP”]
PEEP
Positive End Expiratory Ventilation applies ‘a positive pressure to the patient during the expiratory phase’ to ‘increase the functional residual capacity of the lungs and thus the efficiency of oxygenation of the blood’ (Baker, 2016: 126). ‘In order to be able to deliver a set tidal volume against variable lung compliance and airway resistance, resuscitation ventilators should be time-cycled, volume preset flow generators’ (Baker, 2016: 132).